Quality assurance policies are important for monitoring the consistency and reliability of MRRC equipment. We have several QA policies:
-
Daily Siemens 64-channel head/neck coil
Weekly trigger QA - We test the pulse-to-trigger converter every week to guarantee fMRI experiments will be properly aligned with the scanners.
Developed by: Andrew Reineberg
Performed by: Technologist Team
Method: 1. Open notepad on E-Prime computer, 2. Run sequence named “” in REI-TEST1 protocol, and 3. At end of sequence, use character count in Notepad to ensure 600 triggers were recorded. If fewer than 600 triggers were recorded, immediately inform Chan Moon, Steve Hegedus, and Andrew Reineberg.
Weekly Helium Level Check
Developed by: Adam Cihon and Andrew Reineberg
Method: Review He level on scanner. If less than 45%, call Siemens to ask when Air Products will be refilling the scanner.
Weekly fMRI QA with bullet phantom
Weekly Siemens 32-channel head coil